Mechanism of operation of Automated external defibrillator
An AED is external because the operator applies the electrode pads to the bare chest of the victim, as opposed to internal defibrillators, which have electrodes surgically implanted inside the body of a patient.
Automatic refers to the unit’s ability to autonomously analyse the patient’s condition, and to assist this, the vast majority of units have spoken prompts, and some may also have visual displays to instruct the user.
When turned on or opened, the AED will instruct the user to connect the electrodes (pads) to the patient. Once the pads are attached, everyone should avoid touching the patient so as to avoid false readings by the unit. The pads allow the AED to examine the electrical output from the heart and determine if the patient is in a shockable rhythm (either ventricular fibrillation or ventricular tachycardia). If the device determines that a shock is warranted, it will use the battery to charge its internal capacitor in preparation to deliver the shock. This system is not only safer (charging only when required), but also allows for a faster delivery of the electrical current.
When charged, the device instructs the user to ensure no one is touching the victim and then to press a button to deliver the shock; human intervention is usually required to deliver the shock to the patient in order to avoid the possibility of accidental injury to another person (which can result from a responder or bystander touching the patient at the time of ths shock). Depending on the manufacturer and particular model, after the shock is delivered most devices will analyze the victim and either instruct that CPR be given, or administer another shock.
Many AED units have an ‘event memory’ which store the ECG of the patient along with details of the time the unit was activated and the number and strength of any shocks delivered. Some units also have voice recording abilities[citation needed] to monitor the actions taken by the personnel in order to ascertain if these had any impact on the survival outcome. All this recorded data can be either downloaded to a computer or printed out so that the providing organisation or responsible body is able to see the effectiveness of both CPR and defibrillation.
AEDs available to the public may be semi-automatic or fully automatic. Fully automatic units are likely to have few buttons, often activating as soon as the case is opened, and possibly just one button to shock, or in some cases this will be performed automatically. The user has no input in the operation of the unit apart from attaching the pads and following the prompts. Health care professionals and other trained responders may use a semi-automatic defibrillator, which is likely to have an ECG readout display, and the possibility to override the rhythm analysis software. This allows trained personnel to provide a higher level of care.
The first commercially available AEDs were all of a monophasic type, which gave a high-energy shock, up to 360 to 400 joules depending on the model. This caused increased cardiac injury and in some cases second and third-degree burns around the shock pad sites. Newer AEDs (manufactured after late 2003) have tended to utilise biphasic algorithms which give two sequential lower-energy shocks of 120 - 200 joules, with each shock moving in an opposite polarity between the pads. This lower-energy waveform has proven more effective in clinical tests, as well as offering a reduced rate of complications and reduced recovery time.[10]
Tags: Emergency medical equipment, Medical equipment
This entry was posted
on Monday, June 8th, 2009 at 12:53 am and is filed under Uncategorized.
You can follow any responses to this entry through the RSS 2.0 feed.
You can leave a response, or trackback from your own site.
An automated external defibrillator or AED is a portable electronic device that automatically diagnoses the potentially life threatening cardiac arrhythmias of ventricular fibrillation and ventricular tachycardia in a patient,[1] and is able to treat them through defibrillation, the application of electrical therapy which stops the arrhythmia, allowing the heart to reestablish an effective rhythm.
AEDs are designed to be simple to use for the layman, and the use of AEDs is taught in many first aid, first responder and basic life support (BLS) level CPR classes.[2]
A development of the AICD is a portable external defibrillator that is worn like a vest.[7] The unit monitors the patient 24 hours a day and will automatically deliver a biphasic shock if needed. This device is mainly indicated in patients awaiting an implantable defibrillator. Currently only one company manufactures these and they are of limited availability.
Automated external defibrillators are now easy enough to use that most states in the United States include the "good faith" use of an AED by any person under the Good Samaritan laws.[12] "Good faith" protection under a Good Samaritan law means that a volunteer responder (not acting as a part of one's occupation) cannot be held civilly liable for the harm or death of a victim by providing improper or inadequate care, given that the harm or death was not intentional and the responder was acting within the limits of their training and in good faith. In the United States,
Most manufacturers recommend checking the AED before every period of duty or on a regular basis for fixed units. Some units need to be switched on in order to perform a self check; other models have a self check system built in with a visible indicator.[citation needed]
All manufacturers mark their pads with an expiry date, and it is important to ensure that the pads are in date. This is usually marked on the outside of the pads. Some models are designed to make this date visible through a 'window', although others will require the opening of the case to
Unlike regular defibrillators, an automated external defibrillator requires minimal training to use. It automatically diagnoses the heart rhythm and determines if a shock is needed. Automatic models will administer the shock without the user's command. Semi-automatic models will tell the user that a shock is needed, but the user must tell the machine to do so, usually by pressing a button. In most circumstances, the user cannot override a "no shock" advisory by an AED. Some AEDs may be used on children - those under 55 lbs (25 kg) in weight or under age 8. If a particular model
Defibrillation is the definitive treatment for the life-threatening cardiac arrhythmias, ventricular fibrillation and pulseless ventricular tachycardia. Defibrillation consists of delivering a therapeutic dose of electrical energy to the affected heart with a device called a defibrillator. This depolarizes a critical mass of the heart muscle, terminates the arrhythmia, and allows normal sinus rhythm to be reestablished by the body's natural pacemaker, in the sinoatrial node of the heart.
Defibrillators can be external, transvenous, or implanted, depending on the type of device used or needed. Some external units, known as automated external defibrillators (AEDs), automate the diagnosis of treatable rhythms, meaning that
These units are a compromise between a full manual unit and an automated unit. They are mostly used by pre-hospital care professionals such as paramedics and emergency medical technicians. These units have the automated capabilities of the AED but also feature an ECG display, and a manual override, where the clinician can make their own decision, either before or instead of the computer. Some of these units are also able to act as a pacemaker if the heart rate is too slow (bradycardia) and perform other functions which require a skilled operator.
An automated external defibrillator is used in cases of life threatening cardiac arrhythmias which have led to cardiac arrest. The rhythms the device will treat are usually limited to:
Ventricular fibrillation (shortened to VF or V-Fib)
Pulseless Ventricular tachycardia (shortened to VT or V-Tach)[1]
AEDs, like all defibrillators, are not designed to shock asystole ('flat line' patterns) as this will not have a positive clinical outcome. The asystolic patient only has a chance of survival if, through a combination of CPR and cardiac stimulant drugs, one of the shockable rhythms can be established, which makes it imperative for CPR to be carried
The condition can often be reversed by the electric discharge of direct current from a defibrillator. Although a defibrillator is designed to correct the problem, and its effects can be dramatic, it is not always successful.
The most well-known type of electrode (widely depicted in films and television) is the traditional metal paddle with an insulated (usually plastic) handle. This type must be held in place on the patient's skin while a shock or a series of shocks is delivered. Before the paddle is used, a gel must be applied to the patient's skin, in order to ensure a good connection and to minimize electrical resistance, also called chest impedance (despite the DC discharge). These are generally only found on the manual external units.
Newer types of resuscitation electrodes are designed as an adhesive pad. These are